XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30 14
technician menu.
See page 23 for further details.
Displayed if the active instrument
is not compatible with the
suspension PCB, or if the “SPCB”
menu is not configured correctly.
Configure "SPCB 1-6" in the
technician menu.
disinfection bottle (white handle)
in the service hatch is empty.
“WATER CL” is set correctly
according to the unit
configuration. Consult the
“Technician menu” on page 26.
Replace the white bottle “XO
Water Disinfection” AO-980.
Verify if the bottle is positioned
correctly.
detection PCB is OFF.
LED OFF= Bottle full
LED ON= Bottle empty/missing
Verify connections on the cable
CA-033.
Check that the rubber part on the
sensor is not damaged or
squeezed. Replace if necessary,
More information can be found
“Water Disinfection System” on
page 49 and 50.
Displayed when the Suction
disinfection bottle (yellow handle)
in the service hatch is empty.
“SUCTION CL” is set correctly
according to the unit
configuration. Consult the
“Technician menu” on page 26.
Replace the yellow bottle
“XO Suction Disinfection”.
Follow the instructions explained
in the chapter “CHECK WHITE
“CALL SERVICE”
Displayed when the Suction
disinfection procedure fails. The
error can be caused when the
filling of the mixing cup takes
more than 90 seconds or takes
more than 4 minutes to empty
Check the function of the
“Sensor PCB” located in the
Check the water supply to the
mixing cup.
Check that suction is available
Check that the yellow tubes are
reaching the bottom of the
mixing cup.
Check that the yellow tubes on
the service hatch is not clogged.