XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30 14
instrument.
X” and “SPCB X” in the
technician menu.
See page 23 for further details.
“WRONG SUSP PCB”
Displayed if the active instrument
is not compatible with the
suspension PCB, or if the “SPCB”
menu is not configured correctly.
Configure "SPCB 1-6" in the
technician menu.
“CHECK WHITE BOTTLE”
Displayed when the Water
disinfection bottle (white handle)
in the service hatch is empty.
Check that the parameter
WATER CL” is set correctly
according to the unit
configuration. Consult the
Technician menu” on page 26.
Replace the white bottle “XO
Water Disinfection” AO-980.
If the bottle is full...
Verify if the bottle is positioned
correctly.
Verify LED on the bottle
detection PCB is OFF.
LED OFF= Bottle full
LED ON= Bottle empty/missing
Verify connections on the cable
CA-033.
Check that the rubber part on the
sensor is not damaged or
squeezed. Replace if necessary,
part no. MG-880
More information can be found
Water Disinfection System” on
page 49 and 50.
“CHECK YELLOW BOTTLE”
Displayed when the Suction
disinfection bottle (yellow handle)
in the service hatch is empty.
Check that the parameter
SUCTION CL” is set correctly
according to the unit
configuration. Consult the
Technician menu” on page 26.
Replace the yellow bottle
“XO Suction Disinfection”.
AN-354.
If the bottle is full.
Follow the instructions explained
in the chapter “CHECK WHITE
BOTTLEabove.
“SUCTION CLEAN FAIL”
“CALL SERVICE”
Displayed when the Suction
disinfection procedure fails. The
error can be caused when the
filling of the mixing cup takes
more than 90 seconds or takes
more than 4 minutes to empty
Check the function of the
“Sensor PCBlocated in the
mixing cup. See page 58
Check the water supply to the
mixing cup.
Check that suction is available
Check that the yellow tubes are
reaching the bottom of the
mixing cup.
Check that the yellow tubes on
the service hatch is not clogged.
Consult a XO technician.