Appendix II: Self-diagnosis information
Icons
Self-diagnosis information
pH
Conductivity
Wrong pH buffer solution or the meter recognition of
calibration solution out of range
√
√
Press key when measuring value is not stable during
calibration
√
√
During calibration, the measuring value is not stable for
≥3min.
√
√
Electrode zero electric potential out of range
(<-60mV or >60mV)
√
Electrode slope out of range
(<85% or >110%)
√
pH measuring range out of range
(<-2.00 pH or >16.00pH)
√
Enter in pre-set due calibration to remind due calibration
√
√