Appendix II: Self-diagnosis information
Icons
Self-diagnosis information
pH
Conductivity
Wrong pH buffer solution or the meter recognition of
calibration solution out of range
Press key when measuring value is not stable during
calibration
During calibration, the measuring value is not stable for
≥3min.
Electrode zero electric potential out of range
<-60mV or >60mV
Electrode slope out of range
<85% or >110%
pH measuring range out of range
<-2.00 pH or >16.00pH
Enter in pre-set due calibration to remind due calibration
70+ Series Version 1.0 August 2017