Vertical Pressure Steam Sterilizer
SM520/820
SM530/830
First Edition
Thank you for purchasing SM series vertical pressure steam steriizer of Yamato
Scientific.
To use this unit properly, read this "Instruction Manual" thoroughly before using
this unit.
Keep this instruction manual around this unit for referring at anytime.
WARNING!:
Carefully read and thoroughly understand the important warning items
described in this manual before using this unit.
Yamato Scientific America Inc.

Questions and Answers

  • S
    Sarah ClarkAug 16, 2025
    Why won't the sterilization temperature or pressure rise in my Yamato Laboratory Equipment?
    • G
      Gary BatesAug 17, 2025
      If the sterilization temperature or pressure inside the chamber of your Yamato Laboratory Equipment isn't rising, several factors could be at play. It could be that the preset value is lower than the current temperature inside the chamber, in which case you should adjust the preset value. Also, verify that the power supply voltage is adequate and that the ambient temperature isn't too low. Ensure the cover is securely closed and inspect the packing or flange for any damage.
  • S
    smithkristopherAug 19, 2025
    How to troubleshoot a Yamato Laboratory Equipment that won't start?
    • D
      Diana ReynoldsAug 19, 2025
      If your Yamato Laboratory Equipment isn't starting even with the leakage breaker on, make sure the power cable is properly connected to the power supply. Also, check to see if there is a power outage or if the power voltage is correct.
  • M
    Mr. Nicholas CookAug 22, 2025
    What to do if too much vapor blows out of my Yamato Laboratory Equipment?
    • R
      robinsonronaldAug 22, 2025
      If your Yamato Laboratory Equipment is expelling too much vapor or hot water from the drain bottle, check the cooling water box to ensure it contains water and that the water isn't hot. Make sure the exhaust hose isn't removed or broken, the silicon plug is securely fitted, and the water level in the cooling water box is not above the drain level.
  • E
    Elizabeth WalterAug 25, 2025
    Why does the temperature change during operation of my Yamato Laboratory Equipment?
    • S
      Scott WagnerAug 25, 2025
      If the temperature fluctuates during operation of your Yamato Laboratory Equipment, ensure that an adequate preset temperature is set. Also, check the power supply voltage and the ambient temperature to see if variations are too large.
  • C
    Craig MccoyAug 28, 2025
    What causes exhaust failure in Yamato Laboratory Equipment?
    • J
      Jacob LopezAug 28, 2025
      If you're experiencing exhaust failure or the safety valve is activated on your Yamato Laboratory Equipment, check the hose to the cooling water box for any twists or clogs. Also, inspect the exhaust outlet inside the chamber to ensure it isn't blocked by sterile samples and verify that you haven't stored too many samples.
  • M
    Mary DaltonAug 30, 2025
    What to do if the cover of my Yamato Laboratory Equipment does not open?
    • T
      Toni ColemanAug 30, 2025
      If the cover of your Yamato Laboratory Equipment won't open, first ensure that the power is turned on. If it is, check whether the sterilization process has been completed. If not, wait for it to finish.
  • M
    Ms. Heather GardnerSep 2, 2025
    Why is pressure rising in my Yamato Laboratory Equipment chamber when the solenoid valve is open?
    • C
      Christian LawsonSep 2, 2025
      If the pressure inside the chamber of your Yamato Laboratory Equipment rises even when the solenoid valve is open, check the exhaust outlet inside the chamber to ensure it is not blocked.
  • M
    michael78Sep 4, 2025
    Why does my Yamato Laboratory Equipment halt in standby state?
    • J
      josephmillerSep 4, 2025
      If your Yamato Laboratory Equipment halts in standby state, ensure that the cover is securely closed.
  • M
    michaelharrisSep 7, 2025
    Why is my Yamato Laboratory Equipment leaking water?
    • J
      Jim BrownSep 7, 2025
      If your Yamato Laboratory Equipment is leaking water, check to make sure the drain valve is securely closed.
  • S
    sandragreenSep 10, 2025
    Why is my Yamato Laboratory Equipment making a large noise during air purge?
    • B
      benjamincolleenSep 10, 2025
      If your Yamato Laboratory Equipment is making a lot of noise during air purge, verify that the silencer is installed. Also, check the connection of the silencer inside the cooling water box.

Summary

Safety Precautions

Safety Symbols and Explanations

Explains warning/caution symbols and their meanings.

Critical Safety Warnings

Lists essential warnings to prevent severe hazards.

Cautionary Statements

Details important cautions to avoid minor hazards.

Before Using

Installation Precautions

Covers site selection, environment, and spacing.

Preparation Before Operation

Step-by-step guide to ready the unit for use.

Operation Method

Selecting Operation Modes

Lists and describes available sterilization cycles.

Handling Precautions

General Handling Precautions

Essential safety advice for operating the unit.

Maintenance Method

Daily Inspection and Maintenance Procedures

Routine checks for safe operation.

In the Event of Failure

Failure: Safety Devices and Error Codes

Identifies failure modes and error indicators.

Failure: Troubleshooting Common Issues

Provides solutions for operational problems.

Specification

List of Dangerous Substances

Dangerous Substances List

Lists prohibited hazardous materials.

Regular Spot Check

Yamato SM820 Specifications

General IconGeneral
BrandYamato
ModelSM820
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals