<7. Operation>
7-10
IM 01C25B01-01E
7.6.7 Pressure LRV/URV Conguration
Pressure LRV and URV can be set. The number for each digit is changed by turning the zero-adjustment
screwandsetbypressingthepushbutton.Pleasereferto7.6.8DampingTimeConstantCongurationfor
how to change the numerical value.
When the setting is out of the limit, an alarm will be generated.
7.6.8 Damping Time Constant Conguration
Thedampingtimeconstantfortheamplierassemblycanbeset.QuickResponseParameterisautomatically
set to ON when the damping time constant is set to less than 0.5 seconds. Damping time constant is rounded
ototwodecimalplaces.
F0714.ai
screw
Change the first digit
Set all other
characters in the
same way
Setting range: 0.00 to 100.00 (seconds)
Blinking Blinking Blinking
Save ?/
Cancel ?
push
Go to the next
Character
7.6.9 Output Mode Conguration
Pressure Output Mode (Linear or Sq root) can be selected by turning the zero-adjustment screw. Please refer
to7.6.6PressureUnitCongurationforhowtoselectandsettheenumeratedvalue.
7.6.10 Display Out 1 Conguration
Display Out1 can be selected by turning the zero-adjustment screw. Please refer to 7.6.6 Pressure Unit
Congurationforhowtoselectandsettheenumeratedvalue.