Page #54 background imageLoading...
Page #54 background image
5-7
IM 01F06A00-01E
5. PARAMETER SETUP
(7) Item H : Adjust.
These items are for setting of adjustment.
T050308.EPS
H00
H10
H11
H20
H25
H26
H27
H30
H40
H41
H42
H43
H44
H45
H46
H47
H48
H49
H50
H60
W
W
W
W
W
W
W
W
W
W
W
W
W
W
W
W
W
W
R
%
%
D25
mPa
.
s
Hz
%
Hz
%
Hz
%
Hz
%
Hz
%
2
2
0 to 5
0 to 5
0 to 5
0 to 5
2
0 to 5
2
0 to 5
2
0 to 5
2
0 to 5
2
0.0
0.0
1.0
(0)
1024
1.0
(0)
(0)
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
D
D
D
D
D
D
D
D
D
D
D
D
D
D
D
D
D
D
ADJUST
TRIM 4mA
TRIM 20mA
USER ADJUST
REYNOLDS ADJ
(Indicator and Set only H25 : Active)
DENSITY f
VISCOSITY
EXPANSION FA
FLOW ADJUST
(Indicator and Set only H40 : Active)
FREQUENCY 1
DATA 1
FREQUENCY 2
DATA 2
FREQUENCY 3
DATA 3
FREQUENCY 4
DATA 4
FREQUENCY 5
DATA 5
SELF CHECK
-1.00 to 1.00
-1.00 to 1.00
0.00001 to 32000
NOT ACTIVE (0)
ACTIVE (1)
0.00001 to 32000
0.00001 to 32000
NOT ACTIVE (0)
ACTIVE (1)
NOT ACTIVE (0)
ACTIVE (1)
0 to 32000
-50.00 to 50.00
0 to 32000
-50.00 to 50.00
0 to 32000
-50.00 to 50.00
0 to 32000
-50.00 to 50.00
0 to 32000
-50.00 to 50.00
GOOD
ERROR
Menu H (Adjust)
Trimming 4mA
Trimming 20mA
User Adjust
Reynolds Coefficient
Density at operating condition
Viscosity factor
Expansion correction for compressible Gas
Instrumental Error Adjust
First break-point frequency (f1)
First correcting value (d1)
Second break-point frequency (f2)
Second correcting value (d2)
Third break-point frequency (f3)
Third correcting value (d3)
Fourth break-point frequency (f4)
Fourth correcting value (d4)
Fifth break-point frequency (f5)
Fifth correcting value (d5)
Self-diagnostic message
Item Name R / W Data Range Unit
Decimal
Point
Remark
Initial
value
Disp. U / D
A value in “( )” is the data corresponding to the indicator.
(8) Item J : Test
These items are for test of output.
T050309.EPS
J00
J10
J20
J30
J60
W
W
W
R
%
PPS
0.0
0
(0)
D
D
D
TEST
OUT ANALOG
OUT PULSE
OUT STATUS
SELF CHECK
0.0 to 110.0
0 to 10000
OFF (0)
ON (1)
GOOD
ERROR
Menu J (Test)
Current Output
Pulse Output
Status Output
Self-diagnostic message
Item Name R / W Data Range Unit
1
0
Decimal
Point
Remark
Initial
value
Disp. U / D
A value in “( )” is the data corresponding to the indicator.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DY150 and is the answer not in the manual?

YOKOGAWA DY150 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDY150
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals