<2. Handling Cautions>
9
IM01C25A01-01E
b. FM Intrinsically Safe/Nonincendive for
Fieldbus Type (Except for EJX90A)
EJX/EJA-ESeriespressuretransmitterswithoptional
code/FS15areapplicableforuseinhazardous
locations.
Note1.
• ApplicableStandard:
Class3600:2011,Class3610:2010,
Class3611:2004,Class3810:2005
ANSI/ISA-60079-0:2009,ANSI/ISA-60079-11:2009
NEMA250:1991
• Rating
IntrinsicallySafewithEntity/FISCOfor
ClassI,II,IIIDivision1,
GroupsA,B,C,D,E,F,GT4
ClassIZone0AExiaIICT4
Nonincendiveeldwiring/FNICOfor
ClassI,IIDivision2
GroupsA,B,C,D,F,GT4
ClassIIIDivision1T4
ClassI,Zone2GroupIICT4
• Ambienttemperature:–40to60°C
• Enclosure:Type4X
• ElectricalConnection:
1/2NPTfemale,M20female
Note2. Installation
InstallationshouldbeinaccordancewithControl
DrawingIFM024-A12.