EasyManua.ls Logo

YOKOGAWA EJX110A - Page 10

YOKOGAWA EJX110A
57 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
<2. Handling Cautions>
9
IM01C25A01-01E
b. FM Intrinsically Safe/Nonincendive for
Fieldbus Type (Except for EJX90A)
EJX/EJA-ESeriespressuretransmitterswithoptional
code/FS15areapplicableforuseinhazardous
locations.
Note1.
• ApplicableStandard:
Class3600:2011,Class3610:2010,
Class3611:2004,Class3810:2005
ANSI/ISA-60079-0:2009,ANSI/ISA-60079-11:2009
NEMA250:1991
• Rating
IntrinsicallySafewithEntity/FISCOfor
ClassI,II,IIIDivision1,
GroupsA,B,C,D,E,F,GT4
ClassIZone0AExiaIICT4
Nonincendiveeldwiring/FNICOfor
ClassI,IIDivision2
GroupsA,B,C,D,F,GT4
ClassIIIDivision1T4
ClassI,Zone2GroupIICT4
• Ambienttemperature:–40to60°C
• Enclosure:Type4X
• ElectricalConnection:
1/2NPTfemale,M20female
Note2. Installation
InstallationshouldbeinaccordancewithControl
DrawingIFM024-A12.

Related product manuals