MODBUS Communication
19
TI 12A01F01-62EN
2.3.4 mA output settings (1000-)
Address Parameter
Type Description
mA1 mA2 mA3(Ad) mA4(Ad) mA#: # refers to 1 to 4.
1000 1105 1210 1315 mA# int16 Output(0), Simulate(1), O(2)
1001 1106 1211 1316 mA#
(output)
Process
parameter
int16 pH(0), Temperature(1), ORP(2), rH(3),
Conductivity-TC1(4), Conductivity-
TC2(5), Resistivity-TC1(6), Resistivity-
TC2(7), Concentration-TC1(8),
Concentration-TC2(9), Oxygen(10),
Dierential pH(11), Dierential
ORP(12), Dierential CONDUCT(13),
Dierential RESIST(14), Dierential
DO(15), Average pH(16), Average
ORP(17), Average CONDUCT(18),
Average RESIST(19), Average
DO(20), Ratio(21), Passage(%)(22),
Rejection(%)(23), Deviation(%)(24),
pH calc.(VGB)(25), AI(26), Script1(27),
Script2(28), Script3(29), Script4(30),
Script5(31), Script6(32), Script7(33),
Script8(34)
1002 1107 1212 1317 Process
parameter slot
int16 Sensor1-1(0), Sensor1-2(1),
Sensor1-3(2), Sensor1-4(3),
Sensor2-1(4), Sensor2-2(5),
Sensor2-3(6), Sensor2-4(7),
Converter(8)
1003 1108 1213 1318 Output setup int16 Linear(0), Table(1)
1004, 1005 1109, 1110 1214, 1215 1319, 1320
Linear 0% oat arbitrary
1006, 1007 1111, 1112 1216, 1217 1321, 1322
Linear 100% oat arbitrary
1008, 1009 1113, 1114 1218, 1219 1323, 1324
Sub linear 0% oat arbitrary
1010, 1011 1115, 1116 1220, 1221 1325, 1326
Sub linear
100%
oat arbitrary
1012 1117 1222 1327 Burn int16 O(0), Low(1), High(2)
1013, 1014 1118, 1119 1223, 1224 1328, 1329
Damping time oat 0~3600
1015, 1016 1120, 1121 1225, 1226 1330, 1331
mA#
(simulate)
Simulation % oat -11.25~112.5
1017 1122 1227 1332
mA#
(Congure
Hold)
Last or Fixed int16 Last(0), Fixed(1)
1018, 1019 1123, 1124 1228, 1229 1333, 1334
Fixed value mA oat 2.2~22.0
1020 1125 1230 1335 Hold during
calibration
int16 Disabled(0), Enabled(1)