MODBUS Communication
26
TI 12A01F01-62EN
2.3.9 Display settings Others (3400-)
Address Parameter Type Description
3400 Auto Return int16 Disable(0), 10 min.(1), 60 min.(2)
3401 Luminance int16 10%(0), 20%(1), 30%(2), 40%(3), 50%(4),
60%(5), 70%(6), 80%(7), 90%(8), 100%(9)
3402 Backlight saver int16 Disable(0), 10 min.(1), 30 min.(2), 60 min.
(3)
3403 MONITOR Display int16 O󰀨(0), On(1)
3404 Backlight ash Backlight ash int16 O󰀨(0), On(1)
3405 Backlight ash
behavior
int16 Once (5sec)(0), Continue forever(1)
3406 Alarm NE107 int16 O󰀨(0), On(1)
2.3.10 Advanced settings MODBUS setting (2100-)
Address Parameter Type Description
2100 RS485 setting (RS) MODBUS address int16 1~247
2101 Baud rate int16 9600(0), 38400(1), 115200(2)
2102 Parity int16 Even(0), Odd(1), None(2)
2103 Sensor Address setting
(S)
Address CH1-1 int16 1~247
2104 Address CH1-2 1~248
2105 Address CH1-3
2106 Address CH1-4
2107 Address CH2-1
2108 Address CH2-2
2109 Address CH2-3
2110 Address CH2-4