MODBUS Communication
28
TI 12A01F01-62EN
2.3.12 Advanced settings Ethernet setting (2200-)
Address Parameter Type Description
2200 DHCP enable int16 O󰀨(0), On(1)
2201, 2202 IP Address uint32
2203, 2204 Subnet mask uint32
2205, 2206 Default gateway uint32
CAUTION
Ethernet setting is reected onto the system, after FLXA402 reboots.
2.3.13 Advanced settings Others setting (4000-)
Address Parameter Type Description
4000 Temperature int16 °C(0), °F(1)
4001 Pressure int16 kPa(0), PSI(1)
4002 Date format int16 YYYY/MM/DD(0), MM/DD/YYYY(1), DD/MM/YYYY(2)
2.3.14 MODBUS native (4200-)
Address Parameter Type Description
4200-
4205
cancel MODBUS password MODBUS password function
4300-
4305
set MODBUS password
5000-
5199
User note ASCII free space
10000,
10001
mA1 simulate oat Simulate of mA output mA1 to mA4 from MODBUS
master,
normally 0.0
When entering value at 2.2-22.0[mA] , the value
switches to simulate ouput as below.
0: mA1, 1: mA2, 2: mA3, 4: mA4
This register returns to “0” when the power turns o󰀨.
10002,
10003
mA2 simulate oat
10004,
10005
mA3 simulate oat
10006,
10007
mA4 simulate oat
10008 Contact output simulate
(MODE)
uint16
(bit elds)
MODE: set contact output to simulate by the bit
1: Simulate ON, 0: Simulate O󰀨
VALUE: set contact status to simulate by the bit
1: ON, 0: OFF
bit0: S1, bit1: S2, bit2: S3, bit3: S4, bit4-7: Reserved
This register returns to "0" when the power turns o󰀨.
10009 Contact output simulate
Apr. 19, 2019-00