103
<Default setting of dBm measurement reference resistance>
Set a default value of the reference resistance during dBm measurement mode.
1) Display “db-r” by using the LIGHT key.
The reference value appears on the display.
2)Selectthereferenceresistancebyusingthe▲(RANGE)keyor▼(REL)key.
3)PresstheHOLDkeytosavethesetting.
“SEt” appears and then “db-r.”
Setting values of reference resistance
4, 8, 16, 32, 50, 75, 93, 110, 125, 135, 150, 200, 250, 300, 500, 600, 800, 900, 1000,
1200Ω(defaultvalueis600Ω)
<Default setting of detection method during AC measurement> (TY720 only)
Set a default setting of detection methods during AC measurement.
RMS or MEAN: The default setting is RMS.
1) Display “Ac” by using the LIGHT key.
2)Selectthedetectionmethodbyusingthe▲(RANGE)keyor▼(REL)key.
3)PresstheHOLDkeytosavethesetting.
4) “SEt” appears and then “Ac.”
<Default value of LOGGING interval>
Set a default value of the saving interval during LOGGING mode.
1) Display “LG. int” by using the LIGHT key.
The default setting is 1 sec.
2)Selectthesavingintervalbyusingthe▲(RANGE)keyor▼(REL)key.
3)PresstheHOLDkeytosavethesetting.“SEt”appearsandthen“LG.int.”
Settings of saving interval
1,2,5,10,30,60,600,1800sec