4. Set the height adjustment pins (A) at both
sides of the wheelchair to the required
height.
5. Set the rear side of the both interface
mounting frames to the position that
matches the position of the adjustment pins
(X1 - X4) as follows:
a. Remove the screws (B).
b. Adjust the position of the frames (A)
c. Fasten the screws.
6. Put the seating in its original place.
7. Make sure to lock the interface mounting
frames.
8. Check if the wheelchair is stable in all
available positions for seat lift and tilt.
5.4. Tilt-only interface
5.4.1. Replace the Tilt Actuator (Tilt-only Interface)
Preparation
1. Set the interface in the maximum tilted position for easy access.
2. Prepare the wheelchair for maintenance. See section 5.1.
3. Disconnect the interface cables from the controller.
Procedure
1. Make sure that the interface is kept in the
tilted position. For instance by placing a
support.
2. Disconnect the actuator cable.
3. Remove the circlips (B) from the pivot (C) at
both ends of the actuator (A).
4.
WARNING!
Make sure that the tilt
mechanism is supported
safely.
Remove the old actuator.
5. Put the new actuator into the tilt
mechanism.
6. Connect the actuator cable.
Maintenance tasks & adjustments
Service manual - Alex