EasyManua.ls Logo

YSI 5000 - Page 41

YSI 5000
70 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38
Press [SETUP] to change the SOUR parameters. The following screen shows the default
parameters.
Sample / Total 1 / [ 1]
SOUR
Min. Time
[
1
]
min.
Max. Time
[
15
]
min.
[
*
]
SOUR
@
20
Min. Be
g
innin
g
DO
[
5.00
]
m
g
/L SEND: of
f
Min. Endin
g
DO
[
2.00
]
m
g
/L
Solids Wei
g
ht
[
1.000
]
g
/L
UP DOWN DIGIT NEXT
Use the [UP], [DOWN], [DIGIT] and [NEXT] soft-keys to change the following parameters as
necessary:
Sample / Total: Enter the ratio of sample volume to total volume. For example, if
you are diluting 1 to 10 (1 part sample plus
9 parts dilution water)
then enter the number 10. This would be a 10-fold dilution.
Min. Time: Enter the minimum time (in minutes).
Max. Time: Enter the maximum time (in minutes). When the maximum time is
reached, the test will end.
Min. Beginning DO: Enter the minimum level of DO allowed at the start of the test. If the
DO falls below this level, the test will not start.
Min. Ending DO: Enter the minimum level of DO allowed during the test. If the DO
level falls to this value, the test will end.
Solids Weight: Enter the Total Solids or Volatile Suspended Solids concentration of
the sample in g/L. The maximum acceptable value is 31.999 g/L.
SOUR@20: This applies the calculation for temperature correction to 20°C
according to the Farrell and Bhide equation as follows:
SOUR
20
= SOUR
T
x Θ
(20-T)
Where:
Θ = 1.05 above 20°C
1.07 below 20°C
This calculation is only valid for temperature ranges from 10° - 30°C
SEND: on/off This sets the RS232 status at the beginning of the test.
After the parameters have been set, press [ENTER] to save them and return to the SOUR menu.

Other manuals for YSI 5000