2. Specifications
Oxygen Measurement
Ranges: mg/L: 0.0 to 60.0 mg/L
% air saturation: 0.0 to 600.0% air saturation
mbar: 0 to 1500
Accuracy: mg/L: ±0.1% plus 1 Least Significant Digit (LSD)
% air saturation: ±0.1% plus 1 LSD
mbar: ±
±±
±1% plus 1 LSD
Resolution: mg/L: 0.1%, or 0.01 mg/L , whichever is greater
% air saturation: 0.1% air saturation
mbar: 1 mbar
Temperature Measurement
Range: -5.0 to +50.0°C
Accuracy: ±0.1°C
Resolution: 0.01°C
Barometric Pressure Measurement
Range: 450 to 825 mmHg (600 to 1100 mbar)
Accuracy: ±1% plus 1 LSD within ±10°C ambient temperature from calibration point
Resolution: 1 mmHg
Temperature Compensation
The mg/L mode is automatically temperature-compensated to an accuracy of ±1% of DO
readings between 0 and 5°C, and to an accuracy of ±0.6% of readings between 5 and 45°C.
The % air saturation mode is automatically temperature-compensated to an accuracy of ±0.5% of
calibration values between 0 and 5°C, and to an accuracy of ±0.3% of values between 5 and
45°C.