zebris Medical GmbH FDM Technical Data and Operating Instructions Page 30/36
6.3 Cleaning and disinfection
Cleaning
The platform and accessories are cleaned with a moist cloth while the device is switched
off and the mains plug taken out.
Do not use any aggressive agents to clean the measuring system.
Please be absolutely sure to switch off the device and pull the
mains plug out of the socket before you commence disinfecting
and cleaning.
Disinfection
The platform can be disinfected by wiping over with suitable agents. To clean, wipe the
platform with a cloth soaked in disinfection liquid.
Recommended disinfection agent:
Composition approx. 25% ethanol, 35% Propanol
E.g. Mikrozid Liquid / Schülke & Mayr or similar agents
If you apply disinfection agent be sure to follow the recommendations
given by the manufacturer of the disinfection agent strictly. Especially
consider the rules concerning the commended application time of the
agent.
On no account directly pour disinfection fluids or other liquids
over the FDM platform.
Should any liquid enter the platform it is likely to be damaged.
The fluids required for disinfecting and cleaning must be
stored, prepared and kept ready for use exclusively in the
containers provided, in order to avoid them being mistaken for
other fluids.