2.4.2. Zephyr Hardware Install
The BioHarness Device is powered by a Lithium Polymer cell which is charged via a PC USB
Port. To install the hardware, simply connect to your PC in its charging cradle.
Two notification dialogues will display:
When the hardware is installed, the LED should illuminate as constant pale orange, to indicate
the Device is charging.
If the LED is green, the device may have a completely discharged internal cell (insufficient to
establish USB comms with the PC) – leave in the cradle for a few minutes, remove and replace.
The LED should illuminate pale orange.
There is no charge state indication. 90% charge will be reached in hour from fully discharged.
100% is achieved in 3 hours. The charging automatically circuit prevents over-charging.
Constant pale orange LED
indicates device is charging