NON-INVASIVE BLOOD PRESSURE (NIBP)
10 ZOLL M Series Operator’s Guide - Option Insert
Figure 7 Cuff Inflation Menu
On M Series CCT units, the “Inflation Pressure” field is labeled “Adult Inflation Pressure”,
“Pediatric Inflation Pressure”, or “Neonatal Inflation Pressure” if the Patient Type is set to Adult,
Pediatric, or Neonatal respectively.
2. Press the Next softkey if “Inflation Pressure” is not highlighted.
3. Press the Inc or the Dec softkey to select cuff inflation pressure.
4. Press the Return softkey (twice on M Series CCT units) to return to the NIBP menu.
The cuff inflation pressure is set.
Turning Off
Auto Adjust
You should turn off the Auto Adjust feature in situations where the M Series unit might inflate the cuff
to levels that can cause discomfort or harm to the patient. For example:
Patient is being transported over a very bumpy road.
Patient is moving the monitored arm.
Patient is a small child or infant.
When you turn the Auto Adjust feature off, the M Series unit repeats the same cuff inflation pressure
for every measurement. Because this method does not readjust to the patient’s previously measured
blood pressure, do not use fixed cuff inflation pressures unless you have received the proper training.
To turn off Auto Adjust:
1. Press the Cuff Inflation softkey from the NIBP menu. If the Cuff Inflation softkey is not
displayed, see "Displaying the NIBP Menu" on page 7.
NOTE On M Series CCT units, you access the Cuff Inflation softkey from the NIBP Settings menu. Press
the Settings softkey from the NIBP menu, then press the Cuff Inflation softkey. If the Settings
softkey is not displayed, see "Displaying the NIBP Menu" on page 7.
The M Series unit displays the Cuff Inflation menu.
Figure 8 Cuff Inflation Menu
2. Press the Next softkey until “Auto Adjust” highlights.
3. Press the Inc softkey until “OFF” displays.
Figure 9 Auto Adjust Menu
4. Press the Return softkey (twice on M Series CCT units) to return to NIBP menu.
The M Series unit repeats the selected cuff inflation pressure for every measurement.
Inc
Dec
Prev
Next
Return
Inflation Pressure
Auto Adjust
160 mmHg
160 mmH
g
On
j
g
Inc
Dec
Prev
Next
Return
Inflation Pressure
Auto Adjust
160 mmHg
160 mmH
g
On
j
g
Inc
Dec
Prev
Next
Return
Infl
ation Pressu
r
Auto Adjust
160 mmH
g
Off
g

Other manuals for ZOLL M Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL M Series and is the answer not in the manual?

ZOLL M Series Specifications

General IconGeneral
Pacer Rate30 - 180 ppm
Display TypeLCD
Defibrillator TypeManual and AED
Monitor DisplayHigh-resolution LCD
ECG Monitoring3-lead
PacingDemand
Battery TypeLithium-ion
DisplayHigh-resolution LCD
WaveformBiphasic Truncated Exponential (BTE)
Monitoring ParametersECG, SpO2, EtCO2, NIBP, temperature
ConnectivityData port for data transfer
Pacer TypeTranscutaneous
Pacer Output0 - 140 mA
PrinterIntegrated thermal
Energy Selection2, 3, 5, 7, 10, 15, 20, 30, 50, 70, 100, 120, 150, 200 Joules

Related product manuals