EasyManuals Logo

ZOLL M Series Manual

ZOLL M Series
30 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
NON-INVASIVE BLOOD PRESSURE (NIBP)
18 ZOLL M Series Operator’s Guide - Option Insert
Maintaining the NIBP Option
See the M Series Service Manual for instructions on maintaining the NIBP option. Calibrate the
pressure transducers and perform leak tests yearly.
Performing the Daily Checkout Procedure
Perform the following checkout procedure daily to ensure that the NIBP option is functioning
properly. This daily checkout procedure also ensures that medical personnel maintain familiarity with
the proper use of the NIBP option.
For more specific instructions on how to accomplish each step listed below, refer to the related section
in this manual.
To perform the daily checkout procedure:
1. Select the proper sized cuff.
2. Connect the hose to the M Series unit and to the cuff.
3. Apply the cuff to a human test subject.
4. Display the NIBP menu.
5. Set the cuff inflation pressure (if current setting is not appropriate).
6. Take blood pressure measurement.
7. During cuff inflation, inspect cuff, hoses, and connectors for air leakage (if leak is present, correct,
replace or service item).
8. Ensure no errors display.
9. Verify the measurements values are appropriate.
10. Visually inspect hose, hose connector and cuffs for signs of damage (if damaged, replace).
Cleaning the Hose
Clean the hose when it appears dirty.
To clean the hose:
1. Clean the outside surface of hose with a damp cloth.
2. Disinfect with a mild disinfectant solution.
3. Blow dry air through hose.
The hose is now ready for use.
Cleaning Reusable Cuffs
Clean cuffs when they appear dirty.
To clean reusable cuffs:
1. Clean surface with a damp cloth.
2. Disinfect with a mild disinfectant solution.
3. Dry before use.
The cuffs are now ready for use.

Other manuals for ZOLL M Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL M Series and is the answer not in the manual?

ZOLL M Series Specifications

General IconGeneral
Pacer Rate30 - 180 ppm
Display TypeLCD
Defibrillator TypeManual and AED
Monitor DisplayHigh-resolution LCD
ECG Monitoring3-lead
PacingDemand
Battery TypeLithium-ion
DisplayHigh-resolution LCD
WaveformBiphasic Truncated Exponential (BTE)
Monitoring ParametersECG, SpO2, EtCO2, NIBP, temperature
ConnectivityData port for data transfer
Pacer TypeTranscutaneous
Pacer Output0 - 140 mA
PrinterIntegrated thermal
Energy Selection2, 3, 5, 7, 10, 15, 20, 30, 50, 70, 100, 120, 150, 200 Joules

Related product manuals