NON-INVASIVE BLOOD PRESSURE (NIBP)
22 ZOLL M Series Operator’s Guide - Option Insert
REPEAT NIBP MEAS
alternating with NIBP
SIGNAL WEAK
Weak or no oscillometric signal. Check cuff fit and positioning.
Check hose connection to unit.
Check for kinked hose.
Increase cuff inflation pressure if clini-
cally appropriate.
SYSTEM FAULT Communication problem with the NIBP
module.
Cycle power and retry.
If problem persists, return for service.
False high reading Cuff too small.
Cuff not centered over brachial artery.
Cuff too loose.
Patient’s limb below level of heart.
Select larger cuff.
Readjust cuff.
Reapply cuff or select smaller cuff.
Raise patient’s limb to heart level.
False low reading Cuff too large.
Limb above level of heart.
Select smaller cuff
Lower patient’s limb to heart level.
Message/Symptom Problem User Action

Other manuals for ZOLL M Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL M Series and is the answer not in the manual?

ZOLL M Series Specifications

General IconGeneral
Pacer Rate30 - 180 ppm
Display TypeLCD
Defibrillator TypeManual and AED
Monitor DisplayHigh-resolution LCD
ECG Monitoring3-lead
PacingDemand
Battery TypeLithium-ion
DisplayHigh-resolution LCD
WaveformBiphasic Truncated Exponential (BTE)
Monitoring ParametersECG, SpO2, EtCO2, NIBP, temperature
ConnectivityData port for data transfer
Pacer TypeTranscutaneous
Pacer Output0 - 140 mA
PrinterIntegrated thermal
Energy Selection2, 3, 5, 7, 10, 15, 20, 30, 50, 70, 100, 120, 150, 200 Joules

Related product manuals