NON-INVASIVE BLOOD PRESSURE (NIBP)
24 ZOLL M Series Operator’s Guide - Option Insert
* Neonatal and pediatric NIBP modes available on M Series CCT units only.
Table 11 Battery Operating Time Specifications
For a new, fully charged battery
pack at 20°C with NIBP, EtCO
2
and
SpO
2
Options
35 defibrillator discharges at maximum energy (360J), or 1.25 hours
minimum of continuous ECG monitoring and blood pressure
measurements once every 5 minutes, or 0.75 hour of continuous ECG
monitoring/pacing at 60 mA, 70 beats/min.
Table 12 Environmental Specifications
Operating Temperature 0° to 50° C
Storage and Shipping Temperature -20° to 70° C
Electromagnetic Immunity AAMI DF-2: 1996, EN 61000-4-3: 1996, 20 V/m
Humidity 5 to 95% relative humidity, non-condensing
Operating Pressure 594 to 1060 mBar
Table 13 Alarm Limits Specifications
Systolic Adult High 80 – 260 mmHg;
Low 40 – 140 mmHg;
High 10.7 – 34.7 kPa
Low 5.3 – 18.7 kPa
Pediatric* High 80 – 160 mmHg;
Low 35 – 140 mmHg;
High 10.7 – 21.3 kPa
Low 4.7 – 18.7 kPa
Neonatal* High 60 – 130 mmHg;
Low 25 – 120 mmHg;
High 8.0 – 17.3 kPa
Low 3.3 – 16.0 kPa
Diastolic Adult High 50 – 200 mmHg;
Low 25 – 100 mmHg;
High 6.7 – 26.7 kPa
Low 3.3 – 13.3 kPa
Pediatric* High 50 – 130 mmHg;
Low 20 – 100 mmHg;
High 6.7 – 17.3 kPa
Low 2.7 – 13.3 kPa
Neonatal* High 30 – 105 mmHg;
Low 20 – 100 mmHg;
High 4.0 – 14.0 kPa
Low 2.7 – 13.3 kPa
Mean Adult High 60 – 220 mmHg;
Low 30 – 120 mmHg;
High 8.0 – 29.3 kPa
Low 4.0 – 16.0 kPa
Pediatric* High 60 – 140 mmHg;
Low 20 – 120 mmHg;
High 8.0 – 18.7 kPa
Low 2.7 – 16.0 kPa
Neonatal* High 35 –110 mmHg;
Low 20 – 105 mmHg;
High 4.7 – 14.7 kPa
Low 2.7 – 14.0 kPa

Other manuals for ZOLL M Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL M Series and is the answer not in the manual?

ZOLL M Series Specifications

General IconGeneral
Pacer Rate30 - 180 ppm
Display TypeLCD
Defibrillator TypeManual and AED
Monitor DisplayHigh-resolution LCD
ECG Monitoring3-lead
PacingDemand
Battery TypeLithium-ion
DisplayHigh-resolution LCD
WaveformBiphasic Truncated Exponential (BTE)
Monitoring ParametersECG, SpO2, EtCO2, NIBP, temperature
ConnectivityData port for data transfer
Pacer TypeTranscutaneous
Pacer Output0 - 140 mA
PrinterIntegrated thermal
Energy Selection2, 3, 5, 7, 10, 15, 20, 30, 50, 70, 100, 120, 150, 200 Joules