Page 6
Pacing/Cardioversion
Pacing
1. Place OneStep Pacing or OneStep Complete Electrodes on
patient. To pace with other defibrillation electrodes, they must
be placed on the patient and the three- or five-lead cable
attached to the defibrillator.
2. Turn Selector Switch to green area labeled PACER.
3. Turn the pacer RATE knob to desired rate.
4. Turn the pacer OUTPUT knob clockwise slowly until capture.
5. Slowly decrease output by turning OUTPUT knob counter-
clockwise until capture disappears, then increase by 10%.
6. Check pulse to verify mechanical capture.
7. To check patient’s underlying rhythm; press and hold 4:1 button.
Note: When using BLS or Plus models, manual mode must be activated to pace.
Examples of paced rhythms
Synchronized Cardioversion
1. Apply electrodes or paddles to patient.
2. Turn selector switch to DEFIB. If using BLS or Plus device,
press Manual then Confirm softkeys instead.
3. Press Sync On/Off softkey.
4. Verify that arrows (i) display over R-waves and display reads SYNC.
Do not cardiovert if SYNC is OFF.
5. Follow Manual Defibrillation instructions as described on page 4 and
institution protocol for energy settings.
Dial settings for
R Series Pacing

Other manuals for ZOLL R Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL R Series and is the answer not in the manual?

ZOLL R Series Specifications

General IconGeneral
Defibrillator TypeManual and AED
PacingYes
ECG MonitoringYes
SpO2 MonitoringYes
FAST-SpO2Yes
IP RatingIP55
Defibrillator WaveformBiphasic
Energy Levels1-360 Joules
DisplayLCD
Pacer TypeDemand
Pacer Rate30-180 ppm
Pacer Output0 – 140 mA
EtCO2 MonitoringOptional
NIBP MonitoringOptional
Temperature MonitoringOptional
PrinterIntegrated
Battery TypeLithium-ion
CPR FeedbackYes