55
12.6. Disposing of the component parts
Dispose of or recycle the monitor in an environmentally friendly manner according to
local regulations.
Arm cuff cover
As there is a danger of infection, dispose of the arm cuff cover as medical waste.
Internal backup battery
The monitor is equipped with a lithium battery to back up settings and other data.
Before disposing of the main unit, remove the lithium battery and dispose of it
according to local regulations.
Product name
Model name
Structure name
Material
Package
Box
Cardboard
Packing material
Cardboard
Bag
Vinyl
Inside main unit
Case
ABS/ABS plastic
Internal parts
General parts
Chassis
Steel
Battery on PCB
Lithium battery
Printer unit
Case
ABS/ABS plastic
Internal parts
General parts
Chassis
Steel
External input/output unit
(Option)
Case
ABS/ABS plastic
Internal parts
General parts
12.7. Before requesting service
Before requesting service, please review the following checklist and the error code list
in the next section.
Problem
Check
Countermeasure
Nothing is
displayed when
the power is
switched on.
Is the power cable
connected correctly?
Connect the power cable correctly.
E00
is displayed.
Is there air remaining
in the
cuff?
Wait until the air is released completely
from the cuf
f, and then switch the power on
again.
There is no
pressure.
Is the arm cuff cover pulled
too far over the frames?
See "12.4. Replacing the arm cuff cover" to
reattach the arm cuff cover correctly.