7. Be aware that strong electromagnetic waves can cause malfunctions.
This monitor complies with EMD-standard IEC60601-1-2:2014. However, to
prevent electromagnetic interference with other devices, do not use mobile
phones close to within 30 cm of near the monitor.
If this monitor is located near strong electromagnetic waves, noise may
enter in waveforms and
malfunctions may occur. If unexpected malfunctions
occur during use of this monitor, inspect the electromagnetic environment
and take appropriate actions.
The following are examples of general causes and countermeasures.
Use of mobile phones
Radio waves may cause unexpected malfunctions.
Instruct visitors to rooms or buildings with medical electrical devices not
to use mobile phones or small wireless devices.
High frequency noise is being introduced from other devices via the
electrical outlet.
Check for the source of noise, and then perform countermeasures, such
as using a noise cancellation device on this line.
If the noise source is a device that can be stopped, stop using it.
Use another electrical outlet.
Effects from static electricity are suspected (discharges from devices or the
surrounding area)
Before using the monitor, ensure that the operator and patient have
discharged static electricity.
Humidify the room.
If lightning is occurring nearby, the monitor may receive excessive voltage.
In such cases, power the monitor using the following method.
Use an uninterruptible power supply.
8. Environmental considerations
Before disposing of this monitor, remove the lithium battery from the monitor.