Preparing for Use
For setup instructions on using the i-STAT Alinity v Analyzer and Rechargeable
Battery, Base Station, Printer, and External Simulator, please refer to the Getting
Started Guides enclosed with each item. Contact Abaxis Technical Support if
additional support is needed.
System Specications — i-STAT Alinity v Analyzer
i-STAT Alinity Rechargeable Battery*
Electrical rating:
Operational range-
Temperature and
humidity:
Storage range-
Temperature and
humidity:
Altitude:
3.65 VDC 19.3 Wh (nominal)
10 to 40ºC (50 to 104ºF)
10 to 90% non-condensing relative humidity, with maxi-
mum saturation temperature of 34ºC (93.2ºF)
-10 to 60ºC (14 to 140ºF)
10 to 90% non-condensing, with maximum saturation
temperature of 50ºC (122ºF)
Up to 3,048 meters (10,000 feet)
System Components (Continued)
i-STAT Alinity v — User Manual 650-7100 Rev. A15
*This Lithium-Ion rechargeable battery is designed for use only with the i-STAT Alinity v
Analyzer. Keep a spare, charged battery on hand at all times and store charged battery
in the original packaging.
Operational range -
Temperature and
humidity:
Testing Environment:
Altitude:
Storage range -
Temperature and
humidity:
Power source:
16 to 30ºC (61 to 86ºF) for clinical testing
10 to 40ºC (50 to 104ºF) other than clinical testing
10 to 90% non-condensing relative humidity, with
maximum saturation temperature of 34ºC (93.2ºF)
Indoors, on a dry, clean, horizontal, stable surface.
Avoid nearby vibration equipment such as centrifuges.
Avoid direct sunlight.
Up to 3,048 meters (10,000 feet)
-10 to 60ºC (14 to 140ºF)
10 to 90% non-condensing, with maximum saturation
temperature of 50ºC (122ºF)
Lithium Ion Rechargeable Battery, 3.65 (nominal) VDC,
19.3 Wh