*The VetScan E3+ cartridge containing test for Chloride (Cl) which runs on the VetScan
i-STAT 1 is not currently supported and will not run on the i-STAT Alinity v Analyzer.
Sample Collection and Handling
i-STAT Alinity v — User Manual 650-7100 Rev. A28
Acceptable Sample Types:
Crea CG8+ CG4+ Chem8+ *E3+ EC8+ G 6+
Follow manufacturer’s recommendation for lling the following:
•
Venous whole blood
•
Collect in a lithium heparin tube
•
Collect in a syringe containing balanced heparin
• Arterial whole blood
•
Collect in an arterial blood gas syringe containing balanced heparin
• Capillary whole blood
•
Collect in a balanced heparin capillary tube
•
Recommend 150 μl capillary tube
Not all cartridges are available in all regions. Contact Abaxis Technical Support for
availability in specic markets.
Crea CG8+ CG4+ Chem8+ *E3+ EC8+ G 6+
• Venous and arterial samples
•
Always mix sample immediately before removing for testing
•
Graphics on the analyzer instruct on proper mixing technique
• Capillary samples
•
Fill cartridge immediately after sample collection
Sample Handling