i-STAT Alinity v — User Manual 650-7100 Rev. A37
Best Practice Example
Results
An audible cue will be heard when results are ready.
Touch Silence or remove cartridge to silence the audio.
Results outside of the reference range will be highlighted:
Yellow in the result area indicates that the result is outside of the reference
range, known as an abnormal result. The arrow indicates if the result is high
( ) or low ( ).
Red in the result area indicates that the result is not within the system range.
The arrows indicate if the result is high ( ) or low ( ).
Yellow arrow in page button - indicates one or more results on second
page are outside the reference range.
Red arrow in page button - indicates one or more results on second page
are not within the system range.
White arrow in page button - indicates all results on second page are
within the reference range.
How to Perform Patient Testing (Continued)
Page
Opons
Menu
Na, mmol/L
139 – 150 Ref
129
K, mmol/L
3.4 – 4.9 Ref
3.0
iCa, mmol/L
1.12 — 1.40 Ref
1.2
AnGap, mmol/L
8 – 25 Ref
20
TCO2, mmol/L
17 – 25 Ref
15
Glu, mg/dL
60 – 115 Ref
468
Crea, mg/dL
0.5 – 1.3 Ref
6.2
BUN, mg/dL
10 – 26 Ref
>140
Silence
View
Entered Info
Opons
Menu
Print
Transmit
i-STAT CHEM8+
Pt: 123456
16FEB2025 09:45
Cl, mmol/L
106 – 127 Ref
79