Troubleshooting
i-STAT Alinity v — User Manual 650-7100 Rev. A50
The i-STAT Alinity v Analyzer is programmed to perform quality checks
throughout the testing cycle.
The analyzer has several methods of notifying operators
of failed quality checks.
1. Quality Check Failures
Are displayed when the analyzer identifies a problem while running a
cartridge or simulator
There are 4 types of quality check failures:
1. Analyzer
2. Cartridge
3. Sample
4. Software
Screen displays the type of failure and instructions for resolution
Print
View Resoluon
View
Entered Info
Opons
Menu
Cause
Excess blood added to cartridge.
When filling this cartridge, the blood
advanced past the level indicated by
the 'fill to' arrow.
Help
?
i-STAT CHEM8+
Pt: 654321
Opons
Menu
Home
Cartridge Quality Check Failure
Cartridge Was Overfilled
Code: 37-01
13OCT2025 08:33
Print
View Cause
View
Entered Info
Opons
Menu
Cartridge Quality Check Failure
Resoluon
Cartridge Was Overfilled
When filling a cartridge, use care to
advance blood to the level indicated by
the 'fill to' arrow.
Repeat tesng with a freshly filled
cartridge.
Carefully observe the help provided
throughout the tesng pathway.
If the same quality check failure displays,
contact the system administrator for
further instrucon.
Code: 37-01
Help
?
Opons
Menu
Home
i-STAT CHEM8+
Pt: 654321
13OCT2025 08:33