Troubleshooting
i-STAT Alinity v — User Manual 650-7100 Rev. A50
The i-STAT Alinity v Analyzer is programmed to perform quality checks
throughout the testing cycle.
The analyzer has several methods of notifying operators
of failed quality checks.
1. Quality Check Failures
• Are displayed when the analyzer identifies a problem while running a
cartridge or simulator
• There are 4 types of quality check failures:
1. Analyzer
2. Cartridge
3. Sample
4. Software
• Screen displays the type of failure and instructions for resolution
Print
View Resoluon
View
Entered Info
Opons
Menu
Cause
Excess blood added to cartridge.
When filling this cartridge, the blood
advanced past the level indicated by
the 'fill to' arrow.
Help
?
i-STAT CHEM8+
Pt: 654321
Opons
Menu
Home
Cartridge Quality Check Failure
Cartridge Was Overfilled
Code: 37-01
13OCT2025 08:33
Print
View Cause
View
Entered Info
Opons
Menu
Cartridge Quality Check Failure
Resoluon
Cartridge Was Overfilled
When filling a cartridge, use care to
advance blood to the level indicated by
the 'fill to' arrow.
Repeat tesng with a freshly filled
cartridge.
Carefully observe the help provided
throughout the tesng pathway.
If the same quality check failure displays,
contact the system administrator for
further instrucon.
Code: 37-01
Help
?
Opons
Menu
Home
i-STAT CHEM8+
Pt: 654321
13OCT2025 08:33