Entered Info
Prepare Material and Fill Cartridge
Prepare material per
manufacturer’s
instrucon.
Remove fluid using a syringe with blunt p
needle. Fill cartridge unl sample reaches
fill mark.
Page 1 of 2
Home
i-STAT CHEM8+
i-STAT TriControl L1: 321054
Opons
Menu
View Second Page
Page
Na, mmo/L
129
BUN, mg/dL
119
Glu, mg/dL
468
K, mmol/L
3.0
Crea, mg/dL
6.2
iCa, mmo/L
1.2
Cl, mmo/L
79
AnGap, mmo/L
32
TCO2, mmo/L
19
i-STAT CHEM8+
i-STAT TriControl L1: 321055
Opons
Menu
View
Entered Info
Print
Home
16FEB2025 09:45
i-STAT Alinity v — User Manual 650-7100 Rev. A49
Results
Use Value Assignment Sheet to determine if results are in range. Follow facility
policy if results are outside the assigned range.
For experienced users, the help screens may be bypassed by inserting a lled
cartridge.
Once the cartridge is inserted, Contacting Cartridge will display followed by
the countdown bar. This allows the user to estimate the time to results. Alerts
such as Cartridge Locked and Analyzer Must Remain Level are also displayed.
After the analyzer successfully scans the barcode, help screens will be
displayed.
How to Perform Quality Testing – Liquid Quality Control (Cont.)
How to Perform Quality Testing (Continued)