Scan FLUID LOT
Home
Quality Control Test
View
Entered Info
Next
Previous
Scan CARTRIDGE POUCH Barcode
Home
Quality ControlTest
i-STAT TriControl L1: 321055
i-STAT Alinity v — User Manual 650-7100 Rev. A48
When using i-STAT material, the barcode on the ampule contains the control level
being tested. Barcode on the ampule must be scanned before the ampule
is opened.
Scan FLUID LOT barcode on control vial. Manual entry is not an option.
Scanning is required.
When the text is preceded by a , the information is mandatory.
How to Perform Quality Testing – Liquid Quality Control (Cont.)
How to Perform Quality Testing (Continued)
Scan CARTRIDGE POUCH barcode. Scanning is required. You must scan
the barcode on the cartridge pouch. This information cannot be entered
manually.