ABB MEASUREMENT & ANALYTICS | OPERATING INSTRUCTION
266 with PROFIBUS PA Communication
Pressure transmitters
Engineered solutions for all
applications
Measurement made easy
Introduction
The 2600T family provides comprehensive range of
top quality pressure measurement products,
specifically designed to meet the widest range of
applications ranging from arduous conditions in
offshore oil and gas to the laboratory environment
of the pharmaceutical industry.
For more information
Further publications for 2600T series pressure
products are available for free download from
www.abb.com/pressure
266 models

Table of Contents

Questions and Answers

  • D
    dreedAug 7, 2025
    What to do if ABB 2600T-264 Transmitter detects an overpressure?
    • C
      christopherharrisonAug 8, 2025
      If the ABB Transmitter detects an overpressure because the measured process pressure value is outside the sensor limits and no longer represents the true applied process value, adjust the working range if possible.
  • L
    lesliemurrayAug 12, 2025
    What to do if ABB 2600T-264 shows a pressure overrange?
    • B
      Bonnie JonesAug 13, 2025
      If the ABB Transmitter indicates a pressure overrange because the measured static pressure is above its operational limit, adjust the working range if possible.
  • T
    Tammy GonzalezAug 17, 2025
    How to execute replacement operation on ABB 2600T-264 Transmitter after changing electronics or sensor?
    • C
      Christopher SchmidtAug 17, 2025
      If the electronics or the sensor of your ABB Transmitter have been changed but the replacement operation has not been executed, you must perform the following steps: 1. Move the SW 1 of the electronics to position 1 to enable replace mode. 2. Select the SW 2 to indicate whether the new element is a new sensor or new electronics. 3. Power cycle the device. 4. Move the SW 1 of the electronics to position 0.
  • K
    Katherine WilsonAug 22, 2025
    What to do if ABB 2600T-264 sensor temperature is out of operational limits?
    • D
      Dr. Thomas Anderson MDAug 22, 2025
      If the ABB Transmitter's measured sensor temperature is outside of its operational limits, check the valves and impulse line. If necessary, clean the impulse line and initiate PILD training.
  • K
    Kayla MedinaAug 27, 2025
    What happens if ABB 2600T-264 Transmitter device data is corrupted at start up?
    • R
      rodney97Aug 27, 2025
      If the device data loaded at the start up of your ABB Transmitter are corrupted, precluding the correct functionality of the device because of exchange of data problems between Electronics and Sensor, the device will continue to work without problems, but at the next power cycle the new configuration will be lost.
  • A
    Ashley MartinezSep 1, 2025
    Why is the electronic memory corrupted in my ABB Transmitter?
    • S
      simpsonjoseSep 1, 2025
      If the electronic memory of your ABB Transmitter is corrupted due to a problem in the electronic circuit or in receiver circuit of the sensor, it means that writings to the electronic non-Volatile Memory was not successful.

Summary

Safety Guidelines for Pressure Transmitters

General Safety Information

Overview of safety aspects for device operation and manual adherence.

Safe Use and Personnel Requirements

Covers improper use, technical limits, qualified personnel, and essential safety practices.

Transport, Storage & Installation Safety

Safety for transport, storage, electrical installation, and inspection/maintenance.

Mounting Procedures

General Mounting Principles

Emphasizes careful study of instructions and checking device suitability for the measuring point.

Mounting the Transmitter

Covers factory configuration and hazardous area considerations for mounting.

DP Transmitter Mounting

Guidance on mounting DP transmitters, including bracket use and vertical orientation.

P Style Transmitter Mounting

Guidance on mounting P-style transmitters, emphasizing vertical orientation.

Impulse Piping Connection

Guidelines for laying impulse piping, including length, bends, and filling.

Process Connections

Details on process connection sizes, centers, and flange adapter installations.

Installation Recommendations

Piping configurations for various applications like steam, gas, liquid level, and tank pressure.

Device Introduction and Wiring

Transmitter Wiring

Warnings and guidelines for electrical connections, cable entry, PROFIBUS wiring, and grounding.

PROFIBUS Electronics and Local Controls

On-Board Switches

Details the function of the three on-board switches (SW1, SW2, SW3).

Local Push Button Operations

Explains using Zero, Span, and Write Protection buttons, including SW3 effects.

HMI Local Indicator

Commissioning Procedures

Device Operations

PILD Algorithm for Impulse Line Detection

Maintenance Procedures

Hazardous Area Considerations

Appendices: Device Information and Support

ABB 2600T-264 Specifications

General IconGeneral
TypePressure Transmitter
Model2600T-264
Accuracy±0.075% of span
Operating Temperature-40 to 85°C (-40 to 185°F)
Measurement TypeDifferential Pressure
Output Signal4-20 mA, HART
Power Supply10.5 to 45 V DC
Process Connection1/2-14 NPT, M20 x 1.5
Electrical ConnectionM20x1.5, 1/2-14 NPT
Housing MaterialStainless steel
Diaphragm Material316L SST, Hastelloy C
Enclosure RatingIP66, NEMA 4X