AO2000-LS25 LASER ANALYZERS | OI/AO2000-LS25-EN REV. D 13
Opening and closing the housing
This housing contains a 3 V lithium battery.
Observe the following safety instructions when opening and
closing the housing with the type of protection Ex “p
pressurized enclosure:
DANGER
Explosion hazard
There is a risk of explosion if the housing is opened in a
hazardous atmosphere.
Before opening the housing:
A valid fire permit must be present.
Make sure that there is no explosion hazard.
Before opening the housing, switch off the power supply.
WARNING
Risk of suffocation
Risk of suffocation when using inert purging gases such as
nitrogen (N
2
).
Before opening the housing, switch off the housing
purging.
Recommissioning after opening the housing
Before closing the housing and recommissioning, all dust
deposits in and on the device must be removed.
After closing the housing, the power supply must not be
restored until an initial purge duration of at least 3 minutes
at 20 l/min has elapsed.
Only original spare parts must be used to seal the housing.