Gas Analyzer Calibration 111
Gas Analyzer Calibration
Automatic Calibration: Manual Start
Menu path
Operation Calibration Automatic Calibration
Wait until the Warm-up Phase has Ended
The gas analyzer may not be calibrated until the warm-up phase has
been concluded.
Duration of the warm-up phase:
Uras26: approx. ½ hour without thermostat, approx. 2 hours with
thermostat
Limas23: approx. 2 hours
Magnos206: approx. 1 hour
Magnos27: approx. 2–4 hours
Caldos27: approx. ½ hour
Fidas24: approx. 2 hours
Start Automatic Calibration Manually