4.4.5 Vent
The vent allows the reference air to escape to atmosphere. If the
vent is likely to be exposed to moisture, connect a tube to the
vent and route to a dry area.
Ensure that the vent, or the vent tube, does not become
blocked during probe use.
4.4.6 Reference Air Inlet
A clean, dry, oil-free air supply is required, (for example, from a
pump or filter-regulator – see Appendix D, page 36.
Connect the reference air tubing to the external reference air
inlet – see Fig. 4.8.
4.4.7 Inline Shut-off Valves
On non-AutoCal systems, if the test gas line is connected
permanently and is >10 m (33 ft.) from the supply, fit non-return
valves at the probe head to prevent breathing of flue gas in and
out of test gas lines.
Fig. 4.7 Vent
Note.
The reference air supply must be to instrument air
standards – clean, dry and free of oil vapor and
particle contamination; see schematics in Sections
3.5.2 and 3.5.3 on pages 9 and 10.
ABB recommend using the ABB 5µm / Oil Coalescing
Filter Regulator to supply reference air – see Appendix
D, page 36.
If oil vapor is suspected, use an ABB Reference Air
Pump to supply reference air to the probe – see
Appendix D, page 30.