EasyManua.ls Logo

ABB Endura AZ20 series

ABB Endura AZ20 series
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Endura AZ20 series probe
Combustion oxygen monitor 5 Start-up and Operation
IM/AZ20P–EN Rev. G 27
5.2.2 AutoCal System without Restrictors
To set up an AutoCal system without restrictors:
1. Make test gas and reference air connections as detailed in
Section 3.5.2, page 9 (remote or integral transmitter).
2. Turn on Test Gas 1 and set the pressure to a nominal
1bar (15 psi).
3. At the transmitter, check the Calibrate / AutoCal Hardware
/ Hardware Type parameter is Internal.
4. At the transmitter, open the
'Test Gas 1' valve by selecting:
Calibrate / AutoCal Hardware / Valve Manual Control /
Test Gas 1 and pressing to open the valve.
A small icon is displayed to indicate the valve is in the
Open position:
5. Fine tune Test Gas 1 pressure to 1 bar (15 psi) and set the
flow on the flowmeter to 2.2 l / min (4.662 scfh).
6. Turn gas off at the transmitter by pressing , then turn
Test Gas 1 off at the supply.
7. Repeat steps 2 to 6 for Test Gas 2 (if present).
8. Refer to IM/AZ20E-EN to perform a calibration when
required.
5.2.3 Non-AutoCal System with Restrictors
To set up an Non-AutoCal system with restrictors:
1. Make test gas and reference air connections as detailed in
Section 3.5.3, page 10 (remote or integral transmitter).
2. Turn on Test Gas 1 and adjust the pressure to
1 bar (15 psi).
3. Turn Test Gas 1 off at the supply.
4. Repeat steps 2 to 4 for Test Gas 2 (if present), connect the
Test Gas 2 line to the probe’s external TG1 connection.
5. Refer to IM/AZ20E-EN to perform a calibration when
required.
5.2.4 Non-AutoCal System without Restrictors
To set up an Non-AutoCal system without restrictors:
1. Make test gas and reference air connections as detailed in
Section 3.5.3, page 10 (remote or integral transmitter).
2. Turn on Test Gas 1, adjust the pressure to 1 bar (15 psi)
and the flow on the flowmeter to 2.2 l / min (4.662 scfh).
3. Turn Test Gas 1 off at the supply.
4. Repeat steps 2 to 4 for Test Gas 2 (if present), connecting
the Test Gas 2 line to the probe’s external TG1
connection.
5. Refer to IM/AZ20E-EN to perform a calibration when
required.
Valve icon
'Open' position
Valve icon
'Closed' position
Note. Test Gas 1 and 2 connections are made to the
probe’s external Test Gas 1 (TG1) connection and
must be switched manually – see section 3.5.3, page
10.
Note. Test Gas 1 and 2 connections are made to the
probe’s external Test Gas 1 (TG1) connection and
must be switched manually – see section 3.5.3, page
10.

Table of Contents

Other manuals for ABB Endura AZ20 series

Related product manuals