40 S4 D184B122U02
Submenu/
Parameter
Submenu/
Parameter 
Submenu/
Parameter
Selections
Value Range
Comments
Warning register
After the Warning Register is accessed the 
text "Warning register" is displayed together 
with the Quantity of warnings set. Using 
ENTER again the set warnings with their 
Warning Numbers and descriptive text are 
displayed. Using ENTER again the question 
is displayed if the warning is to be reset or if 
the Help text is to be displayed. Using STEP 
it is possible to display all the set warnings 
with their Warning Number and descriptive 
text. Using ENTER the set errors can be 
reset.
Quantity: 2
3 (set)
ENTER  * Test Mode *
Reset: Enter
Help text: Step
2
Totalizer reset
Display of the number of power outages 
since start-up
Mains interrupt
0
Submenu
Syst. Adjust
Start System Zero
Manual
Auto. adjustment
Zero value check if required (required for 
matching older model flowmeter primaries).
Manual entry, e.g. after a converter ex-
change. 
Automatic adjustment:Valve must be 
closed, liquid in flowmeter primary must be 
at absolute zero flow. Initiate the automatic 
adjustment with ENTER. The limits for the 
zero value are ± 10%. If the value is outside 
of these limits, the adjustment is rejected.
ENTER 0.0 %
System zero
numeric entry
0.0 %
FSM4000 08/2003
view only Display of the software revision. 08/2003 
identifies the issue date.B. 10 = Revision 
level for the Standard and HART-Protocol 
software.
D699G004U01 B.10
FSM4000 PA3.0 view only Only for software PROFIBUS PA
D699G004U02 B.10