D184B122U02 S4 39
Submenu/
Parameter
Submenu/
Parameter 
Submenu/
Parameter
Selections
Value Range
Comments
Contact input
Off
On
This memory is not displayed for PROFIBUS 
PA.
Test the Contact Input. Before a signal is 
applied to the Contact Input, the text "Off" is 
displayed in this menu; after a signal is 
applied the text "On" is displayed (to check 
with a the simulator, use the Totalizer Reset 
key).
Off
Contact output
Off
On
Test the Contact Output. Toggle the value 
"On"/"Off" - the Contact Output condition 
(open/closed) follows the settings selected 
in the parameter "Progr. in/output " - "Con-
tact output".
Off
Display
Test the Display indications. Every point in 
the dot matrix is activated one after the 
other. After all points have been activated 
the failure of a point is obvious.
Test Mode
Off
On
Only for checking the converter using 
Simulator 55XC4000. 
Off
Simulation Mode
Off
On
Turn the flowrate simulation On/Off. To 
check the operation, a flowrate can be 
simulated. The output values should 
correspond to the equivalent of the simula-
ted flowrate value entered. In the lower line 
of the display the message **Q Simulation** 
is shown. After the simulation program is 
completed, the parameter in the menu 
"Simulation Mode" should be set to "Off". 
Off
Simulation Value
numeric entry
-130 % – +130 %
In the process display the flowrate values 
can be simulated by pressing the STEP or 
DATA-Keys. The Current and Pulse Outputs 
follow the simulated flowrate value.
0.0 %
Submenu
Status
After the Error Register is accessed, the text 
"Error register" is displayed together with 
the Quantity of errors set. Using ENTER 
again the set errors with their Error Numbers 
and descriptive text are displayed. Using 
ENTER again the question is displayed if the 
error is to be reset or if the Help text is to be 
displayed. Using STEP it is possible to dis-
play all the set errors with their Error Number 
and descriptive text. Using ENTER the set 
errors can be reset.
Error register
ENTER Quantity 1
C (stored)
ENTER Ext. Data base
Reset: Enter
Help text: Step
0
Empty pipe