1.3 Specifications
Power Requirements See Section 1.2 Model Number Breakdown
Power Consumption Refer to signal converter instruction manual
Flowmeter Characteristics
Meter Size/Flow Capacity
TABLE 1-1. METER CAPACITY
Flow Ranges
0 to value tabulated
Meter Size
METER
CAPACITY*
Minimum Maximum
inch mm gpm gpm lpm gpm lpm
1
⁄
10
3 1.06 0.021 0.08 1.06 4.0
5
⁄
32
4 2.12 0.042 0.16 2.12 8.0
1
⁄
4
6 5.28 0.11 0.40 5.28 20
3
⁄
8
10 11.9 0.24 0.90 11.9 45
1
⁄
2
15 26.4 0.53 2.00 26.4 100
12552.81.064.0052.8 200
1
1
⁄
2
40 158.5 3.17 12.0 158 600
m
3
/h m
3
/h
2 50 264.2 5.28 1.2 264 60
3 80 792.5 15.85 3.6 792 180
4 100 1057 21.14 4.8 1057 240
    * Flow Velocity (ft/s) = (Operating GPM x 32.81)/METER CAPACITY
Span Factory set at specified range between extremes
listed in table above; can be field adjusted.
Rangeability 50:1
Minimum Liquid Conductivity 5 µS/cm
System Accuracy
(At Reference Conditions)
Flowrate Accuracy
Frequency Output < 2% Meter Capacity
≥ 2% Meter Capacity
± 0.01% Meter Capacity
± 0.5% of rate
RFI Protection Class 3-abc-0.5% (30 V/m-27 to 446 MHz) Per
SAMA Standard PMC 33.1-1978.