30  SPIR I T
IT
  F L O W- X   INS T R U C T I O N   M A N U A L   |   I M/ F L O W X - EN 
 
Remote prover IO 
Ball prover / Compact prover 
 Connect the flow pulse signals to the corresponding run 
modules 
 Connect the detector signals to each run module 
 Connect all applicable prove signals (except the detector 
signals) to the remote prover IO module:  
–  prover pressure(s) 
–  prover temperature(s) 
–  prover densitometer 
–  4-way valve commands and statuses 
–  prover commands and statuses 
 
 
Figure 6-38: Single stream flow computers using a common prover 
IO server module.  
Earth ground connection 
The chassis must be connected to a good earth ground to 
prevent human contact with dangerous voltages and to 
minimize the effects of electrical transients.  
 
For the correct operation of the device it is important to 
verify that there is no (significant) difference between 
the voltage levels of the signal ground (0 V of the power supply) 
and the earth ground. 
 
The following pictures show the location of the threaded stud 
for the connection of the earth ground wire for the different 
Spirit
IT
 Flow-X enclosures. 
 
 
Figure 6-39 Flow-X/S earth ground connection 
 
 
Figure 6-40 Flow-X/P earth ground connection