18
Message What It Means What To Do
Incorrect test procedure.
For example, putting
blood on the test strip
before inserting it into
the meter or applying
blood before
and
appear
There may be a problem
with the test strip
Meter error
1. Wait for
and
to appear before applying
blood or control solution.
2. Perform a control solution test using a new
test strip. If the test results are within the range
printed on the test strip vial, retest using blood
and a new test strip.
3. If the control solution result is out of range or the
error reappears, contact Customer Care.
There may be a problem
with the test strip
Meter error
1. Perform a control solution test using a new
test strip. If the test results are within the range
printed on the test strip vial, retest using blood
and a new test strip.
2. If the control solution result is out of range or the
error reappears, contact Customer Care.