EasyManua.ls Logo

Abbott FreeStyle - Troubleshooting Common Issues; Meter Not Entering Test Mode

Abbott FreeStyle
78 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
Troubleshooting
What It Means What To Do
1. The meter
does not
enter test
mode after
inserting a
test strip.
Test strip is inserted
upside down, or bottom
end in, or not fully
inserted into the meter
1. Insert test strip print side up, top end in.
No battery is installed
Battery is installed
incorrectly
1. Install battery with (+) facing up.
Dead battery 1. Replace battery. Reset date and time, if
necessary.
2. If meter still does not enter test mode,
contact Customer Care.
Defective test strip or
meter
Blood or foreign objects
put in the test strip port
1. Contact Customer Care.
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
English

Table of Contents

Other manuals for Abbott FreeStyle

Related product manuals