22
Out of Range Control Solution Results
Repeat the test if control solution results are outside the range printed on
the test strip vial.
Stop using the meter if control solution results are consistently outside the
range printed on the test strip vial. Contact Customer Care.
Causes of out-of-range results include:
Expired or bad control solution
Expired or bad test strip
Testing error
Watered-down control solution
Meter malfunction
Test strip deterioration
Control solution test done outside 59° – 104° F (15° – 40° C)