OPTIS™ Integrated Next Operation and Maintenance Manual 600173796 Rev. A 29
Cleaning and Maintenance
Do not perform cleaning or maintenance on the system during a
The OPTIS Integrated Next accessories should be appropriately
Cleaning
The following OPTIS Integrated Next items require cleaning:
DOC and DOC optical cable
TSC
DOC holster and cables
The DOC, DOC holster, and TSC should be cleaned following the facility’s
standard cleaning schedule, or at least every 30 days under normal use.
1. Turn off (or Standby) all system components with accessible power controls
and unplug the power cables.
2. Clean system surfaces and the keyboard with a dry cloth.
3. Clean the LCD surface of the monitor with a lint-free, non-abrasive cloth.
AUTION: Do not clean the LCD surface with detergents or other cleaning
4. Clean the TSC, DOC, DOC holster, and DOC optical cable with a disinfectant
wipe (1:10 Hypochlorite solution) or Cidex
(Glutaraldehyde 3.4%) and a soft
cloth. Be particularly careful not to stress or sharply bend the DOC optical
cable.
Note:
Though enclosed in a bag during use, the DOC is the system component
most exposed to dirt, fluids, and debris.
The TSC and DOC are not waterproof. Be careful not to use excessive
moisture when cleaning these devices. Wipe only. Do not soak or spray.