Control
Solution Testing
Note: Do not remove the test strip from the monitor or
disturb it during the countdown.
Important: If the countdown does not start:
What It Means:
You might not have applied enough control
solution to the test strip.
What to Do:
Apply a second drop of control solution to the
test strip. Refer to your test strip instructions
for use for the number of seconds you have to
apply a second drop.
If the countdown still does not start, or if the number of
seconds you have to apply a second drop have passed,
discard the test strip, turn off your meter, and try again
with a new test strip.
5
At the end of the countdown:
If the beeper is ON, listen for the
beeper.
The control result shows on the
display window.
Note: If you are doing a β-Ketone control solution test,
KETONE will also show with the result.
The result is stored in your monitors memory as a control
result. In your logbook, record the result as a control result.