Control Solution
Testing
5
To mark the test as a control test, Press
and Release the
button once.
shows on the display window.
Important: If you do not mark the test as a control test,
it will be saved in your monitors memory as one of your
personal blood glucose or blood β-Ketone results. This
could affect your blood glucose averages.
Applying Control Solution to the Test Strip
1
Turn the control solution bottle upside down three to
four times to mix the solution.
2
Remove the cap.
3
Apply a drop of control solution
to the test strip in the area shown
here. The control solution is
drawn into the test strip.
4
Continue to touch the control solution to the test strip
until the monitor begins the test.
The monitor begins the test when:
•You hear the beeper, if the beeper is ON.
•The display window shows the status bar.
•Then the display window shows the countdown.
39
Status bar
Countdown