Performing a Test
13
For an explanation, an example:
Selecting the selection field (A) opens a list of relative water contents according to the following
figure:
An expected water content of 0.01% is selected. When pressing button (B) the permissible
and, if necessary, the optimal water content is calculated.
The result of the calculation is shown in the following figure:
The permissible weight must not be less than 2 gram and not greater than 90 gram. The
recommended sample weight is 10...50 gram. We recommend that you always use a sample
weight near the lower limit of the recommended weight.