Performing a Test
23
2.9
End of the test
At the end of the measurement, a selection menu appears (see (E) 2.8 Measuring period“),
whether the measurement should be saved or discarded. If no selection is made within 5
seconds, the measurement is saved automatically.
After saving, the measurement results are displayed as HTML-Report according to 3.1 HTML-
Report. Then the heater is switched off and the main cooler is adjusted to cool the
HydroTracer to 50 °C stand-by temperature.
To indicate that the HydroTracer is in the cooling phase, a warning field (A) and a temperature
scale (B) are displayed. Using the temperature scale, you can see at what temperature the
HydroTracer is currently in the cooling phase.
Note: Never turn off the power supply of the HydroTracer before the device reaches the
default temperature of 50 °c and the main fan turns off.