55
Troubleshooting Checks
6. Check if you followed the testing steps.
7. Check if you chose the correct control
solution level, either 1 or 2, when you
performed the test.
8. Check if the code number on the meter
display matches the code number on
the test strip container.
9. If you are still unsure of the problem…
Action
Read Chapter 4 “Control Testing” and
test again. If you still have problems, call
Accu-Chek Customer Care.
If you chose the wrong control solution
level, you can still compare the control
result to the range printed on the test
strip container.
If they do not match, insert the correct
code chip into the meter and test again.
Repeat the control test with a new test
strip. If you still have problems, call
Accu-Chek Customer Care.

Other manuals for Accu-Chek Aviva

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Aviva and is the answer not in the manual?

Accu-Chek Aviva Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
CodingNo coding required
Battery LifeApproximately 1000 tests
Memory500 test results
Battery Type1 x 3V lithium (CR2032)
Operating Humidity10% to 90% RH
Meter Weight59 g
Measuring Range10 mg/dL to 600 mg/dL
Automatic Shut OffAfter 90 seconds of inactivity