Lock Basal Rate Profiles
Your pump lets you change your hourly basal rates as needed. You may lock
your basal rate profiles to provide extra protection against accidentally changing
a basal rate during normal operation. When this function is activated, your hourly
basal rates cannot be changed.
Follow the steps below:
1.
Press d to move to the SETUP MENU ADVANCED screen.
Press
f to select.
2. Press d to move to the BASAL RA
TE LOCK screen.
Press
a or s to turn the basal rate lock ON
or OFF.
3. Press f to save and exit.
Chapter 5: User Menus ACCU-CHEK Spirit Insulin Pump
133
Note:
If the basal rate lock is set to ON ( ), the programming of the basal
rate profiles 1, 2, 3, 4, and 5 is blocked. To check if the basal rate lock
is ON, press
d to move to the PROGRAM BASAL RATE screen and
press
f to select. The symbol in the BASAL RATE TOTAL screen
indicates the lock of the basal rate programming. (The symbol appears
in the BASAL RATE TOTAL screen only.) Press
d to view the hourly
basal rates. The
a and s functions are blocked.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek insulin pump and is the answer not in the manual?

Accu-Chek insulin pump Specifications

General IconGeneral
BrandAccu-Chek
Modelinsulin pump
CategoryMedical Equipment
LanguageEnglish

Related product manuals