If the error E6: MECHANICAL ERROR occurs again in spite of these actions,
contact your doctor or healthcare team for an alternate therapy plan. Contact
Pump Support at 1-800-688-4578 for more assistance.
Error E7: ELECTRONIC ERROR
An electronic error was discovered during a routine automatic system check
and the insulin delivery has stopped. You cannot turn off and confirm an error
E7: ELECTRONIC ERROR by pressing
f. You must remove the battery instead.
1. Remove or disconnect the infusion set from your infusion site.
2. Insert a new battery into the pump.
3. Prime the infusion set.
4. Reconnect the infusion set (tubing) a
t your infusion site.
5. Put your pump in RUN mode.
If an error E7: ELECTRONIC ERROR occurs during programming (i.e., basal rate,
time and date) check your settings.
If the error E7: ELECTRONIC ERROR occurs again, contact Pump Support at
1-800-688-4578 for assistance. Contact your doctor or healthcare team if
you need an alternate therapy plan.
Chapter 7: Alerts and Errors ACCU-CHEK Spirit Insulin Pump160