EasyManua.ls Logo

Accu-Chek insulin pump - System Check Start-Up Process

Accu-Chek insulin pump
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. Use the cover to gently push the battery all the way in.
Turn the battery cover clockwise to tighten it.
The battery cover is correctly tightened when
the outside of it is even with the casing. Do not
over-tighten the battery cover.
The pump begins its start-up process.
!
Using AA batteries other than those recommended by
Disetronic Medical Systems may significantly reduce the
battery usage time. Batteries other than those recommended
may leak and corrode the battery contacts within your
insulin pump. Therefore, using batteries not recommended
by Disetronic Medical Systems may void the warranty. Do
not insert old or used batteries into your insulin pump as
this may cause a failure at the insulin pump start-up.
Note:
If your pump was in RUN mode before you removed the battery, an
error E8: POWER INTERRUPT will occur when a new battery is inserted.
Press
f twice to confirm and turn off the error. Check the time and
date in the
STOP screen and correct them, if necessary. See the section
“Setting Time and Date” (page 32) for instructions. Put your pump into
RUN mode if necessary.
Chapter 2: Getting Started ACCU-CHEK Spirit Insulin Pump
27
CAUTION

Table of Contents

Related product manuals