Page #73 background imageLoading...
Page #73 background image
The temperature is
above or below the
proper range for the
system. Move to an
area within the
temperature range
indicated for test strip
use in the test strip
package insert, wait ve minutes, and
repeat the test. Do not arti cially heat or
cool the meter.
The batteries are
almost out of power.
Change the batteries
now.
Blood or control solution
was applied to the test
strip before the ashing
drop symbol appeared
on the display. Discard
the test strip and repeat
the test.
An electronic error
occurred or, in rare
cases, a used test strip
was removed and
reinserted. Turn the
meter off and on, or
take the batteries out
for 20 seconds and
reinsert them. Perform a blood glucose
or control test. If the problem persists,
call Roche.
71
43149_NanoPerf_24MSTR_Ins.indd 7143149_NanoPerf_24MSTR_Ins.indd 71 11/25/08 2:49:09 PM11/25/08 2:49:09 PM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Performa Nano and is the answer not in the manual?

Accu-Chek Performa Nano Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Battery LifeApproximately 1000 tests
Dimensions69 x 43 x 20 mm
DisplayLCD
Memory Capacity500 results
Automatic Power Off90 seconds
Battery1 x 3V lithium battery (CR2032)
Test StripAccu-Chek Performa
Data TransferInfrared (IR)
Storage Conditions-25°C to +70°C
Measurement Range0.6 - 33.3 mmol/L
WeightApprox. 40 g (with battery)
Operating Conditions6°C to 44°C, 10% to 90% RH